1.
Transl Vis Sci Technol ; 12(4): 6, 2023 04 03.
Artigo em Inglês | MEDLINE | ID: mdl-37017957

RESUMO

Purpose: To establish a correlation between phacoemulsification tip normalized driving voltage (NDV) and crystalline lens hardness and use it as an objective measure of lens hardness. The study used a phaco tip equipped with previously validated elongation control adjusting the driving voltage (DV) to produce invariant elongation regardless of resistance. Methods: The laboratory study measured the mean and maximum DV of the phaco tip immersed in glycerol-balanced salt solution and correlated the DV with the kinematic viscosity at 25, 50, and 75 µm tip elongation. The NDV were obtained by dividing the DV in glycerol by the DV in the balanced salt solution. The clinical arm of the study recorded DV of 20 consecutive cataract surgeries. The correlation of mean and maximum NDV to Lens Opacities Classification System (LOCS) III classification, patient's age and effective phaco time were evaluated. Results: The mean and maximum NDV correlated with the kinematic viscosity of the glycerol solution (P < 0.001 in all cases). Mean and maximum NDV during cataract surgery correlated with patients' age, effective phaco time, LOCS III nuclear color, and nuclear opalescence (P < 0.001 in all cases). Conclusions: When a feedback algorithm is running, DV variation strictly correlates with encountered resistance in glycerol solutions and real-life surgery. NDV significantly correlates with the LOCS classification. Future developments might include sensing tips that react to lens hardness in real time. Translational Relevance: The study correlates for the first time phaco tip DV and crystalline lens mechanical properties, establishing an objective and reliable measure of lens hardness. This may lead to smart phaco tips reacting to cataract hardness change in real time and sparing ultrasound dispersion.


Assuntos
Catarata , Cristalino , Facoemulsificação , Humanos , Glicerol , Dureza
2.
Transl Vis Sci Technol ; 5(4): 1, 2016 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-27441099

RESUMO

PURPOSE: To measure the hydraulic resistance (HR) of vitreous cutters equipped with a Regular guillotine Blade (RB) or double edge blade (DEB) at cut rates comprised between 0 and 12,000 cuts per minute (CPM) and compare it with vitreous fragment size. This was an in vitro experimental study; in vivo HR measure and vitreous sampling. METHODS: HR, defined as aspiration pressure/flow rate, was measured in balanced salt solution (BSS; Alcon, Fort Worth, TX) (in vitro) and during pars plana vitrectomy of 20 consecutive patients aged 18 to 65, undergoing macular surgery. HR was recorded at increasing cut rates (500-6000 CPM for the RB and 500-12,000 CPM for the DEB; 5 mL/min flow). Vitreous samples were withdrawn and analyzed with Western and collagen type II and IX immunostaining to evaluate protein size. The main outcome measures were hydraulic resistance (mm Hg/ml/min [±SD]) and optic density for Western blot and immunostaining. RESULTS: RB and DEB showed identical HR in BSS between 0 and 3000 CPM. Above 3000 CPM, RB HR steadily increased, and was significantly higher than DEB HR. Vitreous HR was also similar for the two blades between 0 and 1500 CPM. Above 1500 CPM, RB offered a significantly higher resistance. Western blot and immunostaining of vitreous samples did not yield a significant difference in size, regardless of blade type and cut rate. CONCLUSIONS: DEB is more efficient, offering a lower HR than RB over 1500 CPM in human vitreous. There is no viscosity reduction as a function of cut-rate between 1500 and 12,000 CPM, as HR does not vary. TRANSLATIONAL RELEVANCE: Future vitreous cutters will benefit of a DEB; optimal cut rate needs to be defined, and the simple increase of cut rate does not provide benefits after a certain limit to be assessed.

3.
Retina ; 36(7): 1252-9, 2016 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-26655617

RESUMO

PURPOSE: To characterize the fluidics of axial rotating vitreous cutter probe (RT) compared with the standard guillotine (regular blade), when tested in Balanced Salt Solution (Alcon Laboratories, Forth Worth, TX). METHODS: RT and regular blade (RB) cutter probes connected to the same vitrectomy console used a peristaltic pump. The authors measured instantaneous flow through aspiration tubing proximal to the handpiece, fluid velocity, and acceleration at the port by means of particle image velocimetry. RESULTS: Average flow at aspiration tubing of RT and RB did not vary significantly. Regular blade probes produced higher instantaneous flow fluctuation than RT at any considered cut rate (RB 1,600 6.4 ± 5.3 mL/minute; RB 3,000 11.8 ± 6.3 mL/minute; RT 1,600 0.9 ± 0.7 mL/minute, and RT 3,000 1.8 ± 0.8 mL/minute, respectively. P < 0.001 in all cases). Regular blade also yield significantly higher fluid velocity at cutter port compared with RT (RB 1,600 85.8 ± 70.1 mm/second; RB 3,000 81.6 ± 66.4 mm/second; RT 1,600 71.9 ± 40.3 mm/second; and RT 3,000 32.9 ± 20.8 mm/second. P < 0.001 in all cases). Fluid acceleration at the cutter port was higher when the RB was used (RB 1,600 26.85 ± 30.18 mm/second; RB 3,000 33.76 ± 34.09 mm/second; RT 1,600 24.01 ± 21.94 mm/second; and RT 3,000 16.62 ± 17.87 mm/second. P < 0.001 in all cases). CONCLUSION: RT blade design causes less instantaneous flow fluctuation within the aspiration tubing, and also lower fluid velocity and lower acceleration at the cutter port. Fluidics suggests a safer cutting action and a reduced risk of retinal incarceration.


Assuntos
Hidrodinâmica , Vitrectomia/instrumentação , Corpo Vítreo/cirurgia , Acetatos , Combinação de Medicamentos , Humanos , Minerais , Reologia , Cloreto de Sódio
4.
Invest Ophthalmol Vis Sci ; 55(12): 8497-505, 2014 Dec 02.
Artigo em Inglês | MEDLINE | ID: mdl-25468884

RESUMO

PURPOSE: Unexplained visual field loss after pars plana vitrectomy (PPV) has been reported in up to 14% of all uncomplicated cases with signs varying from visual field defect and disc pallor, to optic atrophy, loss of vision, and phthisis bulbi. Among the postulated pathogenic mechanism is ocular hypoperfusion due to insufficient blood pressure (NBP) and/or elevated IOP, or to their mismatch. The purpose of this study is to assess if, to what extent, and for how long the intraoperative simultaneous variation of IOP and NBP causes mean ocular perfusion pressure (MOPP) to drop below values considered safe, during PPV. METHODS: An IOP sensor placed in the infusion cannula recorded 6 readings per second, while arm systolic and diastolic NBP were taken every 5 minutes throughout surgery and deemed stable in between readings. Supine MOPP was calculated as (115/130) mean arterial pressure--IOP. Surgical monitor video overlay displayed all data in real time and saved them for analysis. RESULTS: Average IOP significantly increased during surgery, while NBP decreased, compared to baseline. As a result, intraoperative MOPP decreased an average 37.1% compared to baseline (range, 13.8%-58.6%; P < 0.05). Of 18 patients, 16 (88.8%) had a significant intraoperative MOPP decrease; 15/18 (83.3%) spent more than 20%, and 5/18 (27.7%) more than 50% of the entire surgery below 30 mm Hg MOPP. Surgical maneuvers, such as phacoemulsification, silicone oil removal, and fluid injection, were associated with significant MOPP decrease, while peeling and vitrectomy were not. CONCLUSIONS: The MOPP may decrease significantly in course of PPV, acutely and for longer time. Surgical maneuvers, including silicone oil removal and combined phacoemulsification, pose a higher risk for MOPP reduction. Discretion should be exercised while administering deep sedation, since it may further lower MOPP through undue blood pressure reduction.


Assuntos
Pressão Sanguínea/fisiologia , Oftalmopatias/cirurgia , Pressão Intraocular/fisiologia , Vitrectomia , Adulto , Idoso , Análise de Variância , Oftalmopatias/fisiopatologia , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Hipertensão Ocular/fisiopatologia , Perfusão , Período Perioperatório , Projetos Piloto , Decúbito Dorsal , Vitrectomia/métodos
5.
Invest Ophthalmol Vis Sci ; 55(12): 8289-94, 2014 Nov 20.
Artigo em Inglês | MEDLINE | ID: mdl-25414180

RESUMO

PURPOSE: We report on instantaneous volumetric flow rate of vitreous cutters measured by means of particle image velocimetry (PIV). METHODS: In an in vitro experimental study, vitreous cutters mounting a regular blade (RB) or modified Twedge blade (TB) engineered for higher flow were connected to a console machine equipped with a double peristaltic and Venturi pump, and immersed in balanced salt solution (BSS). Instantaneous flow was measured on aspiration tubing sections proximal to the cutter hand piece. Measures settings were as follows: (1) regular functioning at 3000 and 6000 cuts per minute (cpm) with 300 mm Hg aspiration with both pumps, (2) aspiration tubing clamped proximal to pump cassette, and (3) aspiration tubing clamped proximal to hand piece, and (4) flow fluctuation as a function of cut rate also was calculated. For main outcome measures, instantaneous volumetric flow rate in mL/min and flow fluctuation measured as the standard deviation of flow rate were measured. RESULTS: Regular functioning shows sinusoidal flow oscillating at cut rate frequency, with amplitude between ±50 mL/min at 3000 cpm and ±35 mL/min at 6000 cpm. The TB always determined a bimodal wave and neither blade nor pump type influenced the sinusoidal pattern of flow. Clamping aspiration tubing zeroes flow, but does not influence fluctuation frequency or amplitude. Clamping at the hand piece determined a significantly higher oscillation. Oscillation amplitude retain a typical resonance pattern with significant changes in function of cut rate and resonance occurs at approximately 4000 cpm. CONCLUSIONS: Cutter blade action determines instantaneous flow rate fluctuation that interferes significantly with cutter suction and hampers a steady suction through cutter port. In a surgical scenario, this translates into a higher risk of inadvertent retinal entrapment and lower predictability of cutter behavior, especially at frequency approaching resonance.


Assuntos
Hidrodinâmica , Microcirurgia/instrumentação , Vitrectomia/instrumentação , Corpo Vítreo/cirurgia , Gravação em Vídeo , Vitrectomia/métodos
7.
Retina ; 34(9): 1896-904, 2014 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-24871998

RESUMO

PURPOSE: To assess the efficacy of novel vitreous cutter blades compared with the regular guillotine by means of particle image velocimetry. Tested blades included a regular blade (RB) and newer designs where a circular (hole blade [HB]) or a slit aperture (slit blade [SB]) had been opened proximal to the cutting edge. METHODS: Twenty-three-gauge probes were immersed in BSS or egg albumen, and high-speed video (1,000 frames per second) was recorded. Duty cycle, flow rate, and acceleration generated by Venturi and peristaltic pumps were measured under cutter settings simulating "low-speed" vitrectomy (1600 cuts per minute, 200 mmHg vacuum) and "high speed" vitrectomy (3000 cuts per minute, 300 mmHg vacuum). RESULTS: The SB and HB had a significantly more favorable duty cycle than that of the RB (P < 0.01) and higher BSS flow rate regardless of the aspiration. The SB flow rate in albumen was significantly higher than that of the HB and RB only over 1,000 cuts per minute using a peristaltic pump and at any cut rate with Venturi pump (P < 0.001). The SB also yielded the lowest fluid acceleration than both the HB and SB (P < 0.001 in all cases). CONCLUSION: The HB and SB proved to be significantly more efficient than the RB, with better duty cycle and higher flow rate. The SB generated significantly less fluid acceleration than any other tested blade, regardless of the fluid viscosity and pump technology. The SB design is significantly more efficient and possibly safer than both HB and RB.


Assuntos
Hidrodinâmica , Vitrectomia/instrumentação , Corpo Vítreo/cirurgia , Acetatos , Combinação de Medicamentos , Desenho de Equipamento , Humanos , Minerais , Ovalbumina , Cloreto de Sódio , Gravação em Vídeo
8.
Retina ; 34(3): 558-67, 2014 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-24013257

RESUMO

PURPOSE: To characterize the fluidics of vitreous cutter port in response to aspiration and blade motion using particle image velocimetry techniques. Diverse surgical scenarios and fluid characteristics were replicated. METHODS: The 23-gauge vitreous cutters were immersed in seeded Balanced Salt Solution (BSS) (Alcon, Forth Worth, TX) or egg albumen, and high-speed video was recorded. Fluid velocity, kinetic energy (KE), and acceleration generated by Venturi and peristaltic pumps were measured in aspiration only (200 and 300 mmHg), low-speed vitrectomy (1,600 cuts per minute; 200 mmHg vacuum), and high-speed vitrectomy (3,000 cuts per minute; 300 mmHg vacuum) modes. RESULTS: The Venturi pump generated significantly higher KE than peristaltic pump in BSS (P < 0.0001 for each pair), and aspiration only yielded significantly higher KE. Cutting activation generated significant acceleration (P < 0.001), and the peristaltic pump produced higher positive and negative acceleration peaks (P < 0.001) than the Venturi pump. In egg albumen, the peristaltic pump generated significantly more KE than the Venturi pump (P < 0.001) and perturbed a much wider area. Acceleration was higher for the peristaltic pump in low-speed mode (P < 0.001), whereas in high-speed modality, the Venturi pump produced the highest acceleration peaks (P < 0.001). CONCLUSION: Pump type and blade motion largely influence velocity, KE, and acceleration. In BSS, the Venturi pump induces higher KE and acceleration, although perturbing fluid less diffusely. In egg albumen, the peristaltic pump perturbed a much wider area and induced a higher KE and acceleration than the Venturi pump, even more so at lower cut rates. As a conclusion, particle image velocimetry allowed precise characterization of fluid velocity in response to cutter activation, suggesting a pragmatic approach to surgical scenarios.


Assuntos
Vitrectomia/instrumentação , Albuminas , Cinética , Microcirurgia/métodos , Modelos Biológicos , Cloreto de Sódio , Sucção , Gravação em Vídeo , Vitrectomia/métodos